                   INSTRUMENT CONTROL PARAMETERS:    Nicotine
                   ------------------------------------------

   C:\MSDCHEM\1\METHODS\SLNIC19SPLIT75.M
      Tue May 26 13:39:28 2015

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  Manual
Mass Spectrometer :  Enabled


 No Sample Prep method has been assigned to this method.

=============================================================================
                                 6890 GC METHOD
=============================================================================

OVEN
   Initial temp:  175 C (On)               Maximum temp:  450 C
   Initial time:  1.00 min                 Equilibration time:  3.00 min
   Ramps:
     #   Rate  Final temp  Final time   CRYO (N2)
     1   5.00      180        0.00         Cryo:  Off
     2  35.00      240        0.00         Cryo fault:  Off
     3    0 (Off)                          Cryo timeout:  120.00 min (Off)
   Post temp:  50 C                        Quick cryo cool:  Off
   Post time:  0.00 min                    Ambient temp:  25 C
   Run time:  3.71 min

FRONT INLET (SPLIT/SPLITLESS)           BACK INLET (UNKNOWN)
   Mode:  Split
   Initial temp:  250 C (On)
   Pressure:  11.3 psi (On)
   Split ratio:  75:1
   Split flow:  127.5 mL/min
   Total flow:  131.7 mL/min
   Gas saver:  On
   Saver flow:  20.0 mL/min
   Saver time:  5.00 min
   Gas type:  Helium

COLUMN 1                                COLUMN 2
   Capillary Column                        (not installed)
   Max temperature:  450 C
   Nominal length:  25.0 m
   Nominal diameter:  320.00 um
   Nominal film thickness:  0.52 um
   Mode:  constant flow
   Initial flow:  1.7 mL/min
   Nominal init pressure:  11.3 psi
   Average velocity:  38 cm/sec
   Inlet:  Front Inlet
   Outlet:  (other)
   Outlet pressure:  ambient

FRONT DETECTOR (NO DET)                 BACK DETECTOR (NO DET)

SIGNAL 1                                SIGNAL 2
   Save Data:  Off                         Save Data:  Off

THERMAL AUX 2
   Use:  MSD Transfer Line Heater
   Initial temp:  280 C (On)

                                        POST RUN
                                           Post Time: 0.00 min


TIME TABLE
   Time(min)      Parameter & Setpoint



 Column 1 Inventory Number :  
 Column 2 Inventory Number :  

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : SIM


MS Information
-- -----------

Solvent Delay            : 0.65 min

EMV Mode                 : Gain Factor
Gain Factor              : 1.00
Resulting EM Voltage     : 2424

[Sim Parameters]

GROUP 1
Group ID                 : 1
Resolution               : High
Plot 1 Ion               : 102.00
Ions/Dwell In Group      (  Mass,  Dwell) (  Mass,  Dwell) 
                         (102.00,     10) (129.00,     10) 

GROUP 2
Group ID                 : 2
Resolution               : High
Group Start Time         : 1.40
Plot 1 Ion               : 133.00
Ions/Dwell In Group      (  Mass,  Dwell) (  Mass,  Dwell) (  Mass,  Dwell) 
                         (133.00,     10) (161.00,     10) (162.00,     35) 

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US10490173
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      16.720
 IONFOCUS    :      88.259
 ENTRANCE_LE :      26.000
 EMVOLTS     :    2752.941
                               Actual EMV  :    2423.53
                               GAIN FACTOR :       0.98
 AMUGAIN     :    2267.000
 AMUOFFSET   :     130.000
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      16.314
 MASSGAIN    :     104.000   
 MASSOFFSET  :      -9.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------